Descripción
Innovational Portability and Flexibility in X-ray Room and Beyond
High efficiency power system using high frequency inverter method.
Arm free stop structure without lock system and gas shock absorbers enables free positioning for any directions of desired shooting points.
Inverter method HFG
Inverter method HFG enables efficient and stable X-ray output.
Arm free stop structure
After relocation of head parts where x-ray lights are generated in a direction of top to bottom, side to side for desired shooting points and simply release it to stop. The operation dose not require extra lock and smoothly operates and stops at a desired point to film without vexatious to level lock systems used before.
Embedded break structure
Application of an internal break structure to ensure safety and exterior design because outer mounted products can cause risk of injuries for operating personnel, also prone to dirt increasing unattractive appearance.
Easy operation
Up to 14 APR(Anatomical Programmed Radiography) set-up options with easy operations and ensures convenience of choosing desired images while filming.
H-V Generator
- High Frequency Inverter Control System
Output
- Adjustable Tube Voltage: 40kV~110kV (1kV step)
- mAs Range: 1mAs ~ 200mAs
- Max output power: 5.1k @60kV.85mA / single phase AC 220V
- Max kV/mA: 110kV / 100mA
- Deviation: kV ± 5% , mAs ± 10%
- Waiting time: Min. Exposure interval: 10sec(not continue)
- Duty Cycle: 1/15 duty
X-Ray tube
- Maker: CEI
- Model: OX/110-15
- Focal spot: 1/8mm
- Max. Tube Voltage/Current: 110kV/100mA
- Anode Type: Stationary Anode
- Anode Heat Storage Capacity: More than 20KJ
- Anode Cooling Rate: 270W/sec
- Inherent Filtration: Min. 0.5mmAI (Total Filtration: AL2.5mm)
- Target Angle: 15°
Collimator
- Type: Mobile Type
- Max. Radiation Field: 35cm X 35cm @ 65cm SID
- Min. Radiation Field: 5cm X 5cm @ 1m SID
- Lamp illumination: More than 160 LUX (24V/150W)
- Timer for Lamp: More than 30sec
- SID Scld: 2m
AC Input
- Voltage: AC110/120V, 220/230V
- Phase & Frequency Single phase 50/60Hz
- Allowable Input Volt Deviation: AC Input ± 10%
- Power Capacity (Max): 7.5kVA
For more information about the product and / or quote.
Contact us with the following form.